3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
:SYST:COMM:SOUR:ADDR “USB0::12212::32145::US1234567A::INSTR”
:SYST:COMM:SOUR:ADDR “GPIB1::19::INSTR”
Notes Empty string is allowed and means no source is defined or selected
The address string is the VISA address for external sources and INTERNAL for an internal source
Dependencies Operation with a source requires a license. If the proper license is not installed, the command
generates an error message, “Settings conflict; option not installed”
If no supported source, or no source at all, is found at the specified address, the command generates
an error message
Preset The current source selection is unaffected by Mode Preset and Source Preset but reverts to [None] on
Restore Input/Output Defaults
If an internal Tracking Generator is installed, then instead of None, the default selection is INTERNAL
State Saved Selected Source is saved
Power On Persistent (survives power cycle)
Part of the Input/Output system, which means it is Loaded and Saved with state
Backwards
Compatibility SCPI
:SYSTem:COMMunicate:LAN:SOURce[:EXTernal]:IP <address string>
:SYSTem:COMMunicate:LAN:SOURce[:EXTernal]:IP?
Backwards
Compatibility
Notes
Provided for compatibility with PSA Option215. The address string is reformatted for the X-Series. For
example, if you send:
:SYSTem:COMMunicate:LAN:SOURce:EXTernal:IP 146.208.172.111
The instrument turns this into
:SYSTem:COMMunicate:SOURce:ADDRess “TCPIP0::146.208.172.111::INSTR"
Delete Highlighted Source
Deletes the highlighted source from the list of available sources. You are prompted
with a dialog box to make sure you really want to do this. The prompt says:
“The highlighted source will be permanently deleted from the list. Are you sure you
want to do this? Press Enter to proceed, or Cancel (ESC) to cancel”
Verify Connection
Verifies the interface connection to the Current Source (it does not verify any signal
connections!)
Until the selected source is verified, a statement appears at the bottom of the
screen which says (in red):
This Source has not been verified. Press “Verify Source” to check the interface
connection.
When you press this key, the connection is checked to the selected source. If all is
well, the statement is changed to (in green):
Spectrum Analyzer Mode User's &Programmer's Reference 477