3 Spectrum Analyzer Mode
3.10 TOI Measurement
wideband signals with occupied bandwidths up to 2GHz. This increases the risk
of images failing to be rejected, but improves the measurement speed
Remote Command
[:SENSe]:POWer[:RF]:SWPResel:BW NORMal | NARRow
[:SENSe]:POWer[:RF]:SWPResel:BW?
Example
:POW:SWPR:BW NARR
Dependencies Only appears in N9041B and N9042B+V3050A. Only appears in measurements that use the Swept
method
Grayed-out when "Software Preselection" on page 1992 is OFF. The grayout message is “Unavailable
unless SW Presel enabled”
For N9042B+V3050A, the parameter is SCPI-only, and always set to NARRow when Software
Preselection is enabled
Preset
N9041B
NORMal
N9042B+V3050A
NARRow
State Saved Saved in instrument state
High Freq Prefilter
Lets you set the state of Prefilter for center frequencies above 1310MHz.
In VXT Models M9410A/11A and M9410E/11E in bypass frequency range
(1310MHz~5GHz), the Receiver RF path has a bank of filters that come after the RF
Attenuator in the signal path. Since this bank of filters precedes the mixer they are
known as the “Prefilter” bank. Their purpose is to eliminate unwanted in-band
mixing products by filtering out all but the desired frequencies before the signals get
to the mixer. There are 13 prefilter bands to cover the frequencies between 9kHz
and 6000MHz. The Prefilter provides the necessary rejection of the unwanted
signal.
Remote Command
[:SENSe]:<measurement>:PFILter[:STATe] ON | OFF | 1 | 0
[:SENSe]:<measurement>:PFILter[:STATe]?
Example Enable High Freq Prefilter for the Complex Spectrum Measurement in BASIC Mode:
:SPEC:PFIL ON
Enable High Freq Prefilter for the IQ Waveform Measurement, in multiple Modes:
:WAV:PFIL ON
Enable High Freq Prefilter for the Swept SA Measurement in SA Mode:
:SAN:PFIL ON
Dependencies Only appears in VXT models M9410A/11A with center frequency above 1310MHz, and M9410E/11E
in frequency range 1310MHz~5GHz
Preset See "Prefilter Presets" on page 1805 below
1804 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals