3 Spectrum Analyzer Mode
3.13 Phase and Amplitude vs Time Measurement
Notes If :INIT:CONT is ON, then a new continuous measurement will start immediately, with sweep (data
acquisition) occurring once the trigger condition has been met
If :INIT:CONT is OFF, then :INIT:IMM is used to start a single measurement, with sweep (data
acquisition) occurring once the trigger condition has been met
Dependencies For continuous measurement, :ABORt is equivalent to the Restart key
Not all measurements support this command
Status Bits/OPC
dependencies
The STATus:OPERation register bits 0 through 8 are cleared , except bit 6 (Waiting for Periodic
Sync). Initiating a measurement and Waiting for Periodic Sync could be asynchronous
The STATus:QUEStionable register bit 9 (INTegrity sum) is cleared
Since all the bits that feed into OPC are cleared by :ABORt, the Abort command will cause the *OPC
query to return true
3.13.10 Trace
Trace functionality is not supported in the Phase and Amplitude vs Time
measurement.
Spectrum Analyzer Mode User's &Programmer's Reference 2043