3 Spectrum Analyzer Mode
3.3 Channel Power Measurement
Avg/Hold Number
Specifies the number of measurement averages used to calculate the measurement
result. The average is displayed at the end of each sweep. After the specified
number of average counts, the averaging mode (terminal control) setting
determines the averaging action.
Remote
Command
[:SENSe]:CHPower:AVERage:COUNt <integer>
[:SENSe]:CHPower:AVERage:COUNt?
Example
:CHP:AVER:COUN 15
:CHP:AVER:COUN?
Preset SA, WLAN: 10
WCDMA, LTEAFDD, LTEATDD, 5G NR, MSR: 200
State Saved Saved in instrument state
Min/Max 1 / 10000
Annotation The average count is displayed in the measurement bar on the front panel display. The annotation
appears in the format n/N where n is the current average and N is the average count
Averaging On/Off
Turns averaging on or off for this measurement.
NOTE
In this measurement, the Average Type is preset to the Log-Pwr Avg (Video)
method. Other averaging methods are not available.
Remote Command
[:SENSe]:CHPower:AVERage[:STATe] ON | OFF | 1 | 0
[:SENSe]:CHPower:AVERage[:STATe]?
Example
:CHP:AVER ON
:CHP:AVER?
Preset
ON
State Saved Yes
Range
ON|OFF
Avg Mode
Allows you to select the type of termination control used for the averaging function.
This determines the averaging action after the specified number of data acquisitions
(average count) is reached. Options are:
640 Spectrum Analyzer Mode User's &Programmer's Reference