9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Questionable Calibration Extended Needed Event Query
Returns the decimal value of the sum of the bits in the Questionable Calibration
Extended Needed Event register.
NOTE
The register requires that the associated PTRansition or NTRansition filters be
set before a condition register bit can set a bit in the event register. The data in
this register is latched until it is queried. Once queried, the register is cleared.
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed[:EVENt]?
Example
:STAT:QUES:CAL:EXT:NEED?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Questionable Calibration Extended Needed Negative Transition
Determines which bits in the "Questionable Calibration Extended Needed Condition"
on page 2811 register will set the corresponding bit in the "Questionable Calibration
Extended Needed Event Query" on page 2812 register when the condition register
bit has a negative transition (1 to 0).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:NTRansition <integer>
:STATus:QUEStionable:CALibration:EXTended:NEEDed:NTRansition?
Example
:STAT:QUES:CAL:EXT:NEED:NTR 2
Conducted alignment required bit being cleared will be reported
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Questionable Calibration Extended Needed Positive Transition
Determines which bits in the "Questionable Calibration Extended Needed Condition"
on page 2811 register will set the corresponding bit in the "Questionable Calibration
Spectrum Analyzer Mode User's &Programmer's Reference 2812

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)