7 Save/Recall/Print
7.3 Save
Periodic Timer Sync Source None
Periodic Timer Trigger Delay 1.00E-06
Periodic Timer Trigger Delay
State
FALSE
Preselector Adjust 0
Ref Trace FALSE
RFBurst Trigger Delay 1.00E-06
RFBurst Trigger Delay State FALSE
RFBurst Trigger Level Abs -20
RFBurst Trigger Level Rel -6
RFBurst Trigger Level Type Absolute
RFBurst Trigger Slope Positive
Scale/Div 2
Trigger Holdoff 0.1
Trigger Holdoff State FALSE
TriggerSource Free
MeasResult1 MeasResult2 MeasResult3 MeasResult4
-73.0651058869747 36.9712197125257 36.7879441171442
36.9712197125257 36.8850431211499 36.7032368203129
TXP Meas Results File Contents
A TXP Meas Results File contains measurement results with the following header
information:
–
File ID string, which is “MeasResult”
–
Measurement ID following Mode ID, which is “SA:TXP” for example
–
Firmware rev and model number
–
Option string
–
Center Frequency
–
TXP_InfoBw
–
TXP_Used_CaptureTime
The header data is followed by a line containing entries "MeasResult1" to
"MeasResult4", which flags the start of the measurement results. Each line of
Measurement Results consists of four comma separated-values, from the
MeasResult1 value to the MeasResult4 value.
Spectrum Analyzer Mode User's &Programmer's Reference 2608