3 Spectrum Analyzer Mode
3.10 TOI Measurement
Compatibility SCPI
Span Presets
The following table provides the Span Presets for the Spectrum Analyzer mode, and
the Max Span, for the various frequency options:
Freq Option Span after Mode Preset Max Span
(can't set higher than this)
503 (all but CXA) 3.59 GHz 3.7 GHz
503 (CXA), F03 (CXA-m) 2.99 GHz 3.08 GHz
507 (all but CXA) 6.99 GHz 7.1 GHz
507 (CXA), F07 (CXA-m) 7.49 GHz 7.575 GHz
508 (all but MXE) 8.39 GHz 8.5 GHz
508 (MXE) 3.59 GHz 8.5 GHz
513, F13 13.59 GHz 13.8 GHz
526 (all but CXA and MXE) 26.49 GHz 27.0 GHz
526 (MXE) 3.59 GHz 27.0 GHz
526 (CXA), F26 (CXA-m) 26.49 GHz 26.55 GHz
543 42.99 GHz
544 43.99 GHz 44.5 GHz
550 49.99 GHz 52 GHz
Input 2:
Model Span after Mode Preset Max Span
(can't set higher than this)
CXA opt C75 1.499 GHz 1.58 GHz
MXE 1 GHz 1.000025 GHz
Note that if you are in External Mixing, the maximum Span is Maximum Stop
Frequency – Minimum Start Frequency for the currently selected mixer.
Auto Tune
This is an immediate action control. When it is pressed, it causes the instrument to
change "Center Frequency" on page 1818 to the strongest signal in the tunable
span of the instrument, excluding the LO. It is designed to quickly get you to the
most likely signal(s) of interest, with no signal analysis knowledge required. As such,
there are no configurable parameters for this feature. There are only preselected
values that work in most real-world situations.
1822 Spectrum Analyzer Mode User's &Programmer's Reference