3 Spectrum Analyzer Mode
3.5 ACP Measurement
Backwards Compatibility SCPI Commands
Sets or queries the state of a marker. Setting a marker that is OFF to ON (1) puts it in
POSition mode and places it at the center of the screen.
Example
:CALC:ACP:MARK2:STAT ON
:CALC:ACP:MARK2:STAT?
Preset
OFF
State Saved Saved in instrument state
Range
OFF|ON
Backwards
Compatibility
SCPI
:CALCulate:ACPower:MARKer[1]|2|…|12:STATe OFF | ON | 0 | 1
:CALCulate:ACPower:MARKer[1]|2|…|12:STATe?
Delta Marker (Reset Delta)
Pressing this button has exactly the same effect as pressing DELTa in Marker Mode.
The selected marker becomes a Delta marker. If the selected marker is already a
Delta marker, the reference marker is moved to the current position of the selected
marker, thus resetting the delta to zero.
Marker Settings Diagram
Lets you configure the Marker system using a visual utility.
All Markers Off
Turns off all markers.
Remote Command
:CALCulate:ACPower:MARKer:AOFF
Example
:CALC:ACP:MARK:AOFF
Dependencies Unavailable when "Meas Method" on page 970 is RBW
Couple Markers
When this function is ON, moving any marker causes an equal X-Axis movement of
every other marker that is not OFF. By “equal X-Axis movement” we mean that we
preserve the difference between each marker’s X-Axis value (in the fundamental X-
Axis units of the trace that marker is on), and the X-Axis value of the marker being
moved (in the same fundamental X-Axis units).
Spectrum Analyzer Mode User's &Programmer's Reference 963