9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Bit Condition Operation
0 ODI Connection Error ODI Connection Error
This bit is never triggered, only its aliases are
Filter Registers
"Questionable Integrity Output Condition" on page 2827
"Questionable Integrity Output Enable" on page 2827
"Questionable Integrity Output Event Query" on page 2828
"Questionable Integrity Output Negative Transition" on page 2828
"Questionable Integrity Output Positive Transition" on page 2829
Questionable Integrity Output Condition
Returns the decimal value of the sum of the bits in the Questionable Integrity Output
Condition register.
NOTE
The data in this register is continuously updated and reflects the current
conditions.
Remote
Command
:STATus:QUEStionable:INTegrity:OUTPut:CONDition?
Example
:STAT:QUES:INT:OUTP:COND?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Questionable Integrity Output Enable
Determines which bits in the "Questionable Integrity Output Condition" on page
2827 register will set the corresponding bit in the "Questionable Integrity Output
Event Query" on page 2828 register, which also sets the Data Output Summary bit
(bit 2) in the "Questionable Integrity Register" on page 2819.
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:INTegrity:OUTPut:ENABle
:STATus:QUEStionable:INTegrity:OUTPut:ENABle?
Example
:STAT:QUES:INT:OUTP:ENAB 1
2827 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)