3 Spectrum Analyzer Mode
3.11 Harmonics Measurement
–
The average/hold count k is set to 1, so that the next time the min hold trace is
displayed it simply represents one trace of new data
–
A new sweep is initiated
Remember that restarting Min Hold also restarts Max Hold and averaging, because
there is only one count for Trace Average and Hold.
Clear and Write | Restart Averaging | Restart Max/Min Hold
Starts the trace writing, as though the "Trace Type" on page 1927 had just been
selected. The effect is exactly the same as reselecting the current Trace Type again
– the control is provided because it may not be obvious that reselecting the same
selection from a radio button menu will take any action.
This control displays different labels, depending on the selected Trace Type:
–
Clear/Write: Clear and Write
–
Trace Average: Restart Averaging
–
Max Hold: Restart Max Hold
–
Min Hold: Restart Min Hold
1932 Spectrum Analyzer Mode User's &Programmer's Reference