6 Input/Output
6.1 RF Source
on page 2239. As a step is set up, the value entered runs through several levels of
validation.
Remote Command
:SOURce:LIST:SETup:RADio:BAND <enum>, <enum>, <enum>, …
:SOURce:LIST:SETup:RADio:BAND?
Example
:SOUR:LIST:SET:RAD:BAND PGSM, EGSM, RGSM
:SOUR:LIST:SET:RAD:BAND?
Notes The command sets up the parameter array of whole list sequence
Radio Band <enum> - specifies the radio band for the step. For available options, see "Band" on page
2242
If the input parameter number exceeds the step number defined by "Number of Steps" on page 2239,
then error -221, "Settings conflict; The number of input parameters is too large and is truncated to
current list step number" is generated, and only those parameters whose index number falls within the
number of steps will be updated
Dependencies The range is 1 to 1000, which is determined by the number of steps you have configured. For details
see "Number of Steps" on page 2239
Step Configuration of Radio Band Link parameter list (Remote Command Only)
Configures the Radio Band Link parameter array of the whole List Sequencer at one
time. The number of arrays is same as step number defined in "Number of Steps" on
page 2239. As a step is set up, the value entered runs through several levels of
validation.
Remote Command
:SOURce:LIST:SETup:RADio:BAND:LINK <enum>, <enum>, <enum>, …
:SOURce:LIST:SETup:RADio:BAND:LINK?
Example
:SOUR:LIST:SET:RAD:BAND:LINK DOWN,UP,UP
:SOUR:LIST:SET:RAD:BAND:LINK?
Notes The command sets up the parameter array of whole list sequence
Radio Band Link <enum> - specifies the radio band link direction for the step. Options are:
DOWN|UP
If input parameter number exceeds the step number defined by "Number of Steps" on page 2239,
then error -221, "Settings conflict; The number of input parameters is too large and is truncated to
current list step number" is generated, and only those parameters whose index number falls within the
number of steps will be updated
Dependencies The range is 1 to 1000, which is determined by the number of steps you have configured. For details
see "Number of Steps" on page 2239
2256 Spectrum Analyzer Mode User's &Programmer's Reference