3 Spectrum Analyzer Mode
3.5 ACP Measurement
3.5.13.7 Offset Filter Alpha (Backward Compatibility SCPI)
Example
:ACP:FILT:ALPH 0.5
:ACP:FILT:ALPH?
Couplings This command is an alias of:
[:SENSe]:ACPower:OFFSet[1]|2:LIST:FILTer:ALPHa
Sending the command sets values of all offsets for BS and MS, but the query always returns a value of
BS Offset A
Preset 0.22
State Saved Saved in instrument state
Min/Max 0.01/1.00
Backwards
Compatibility
SCPI
[:SENSe]:ACPower:FILTer[:RRC]:ALPHa <real>
[:SENSe]:ACPower:FILTer[:RRC]:ALPHa?
[:SENSe]:ACPR:FILTer[:RRC]:ALPHa
[:SENSe]:MCPower:FILTer[:RRC]:ALPHa
3.5.13.8 Method for Carrier (Backward Compatibility SCPI)
Example
:ACP:CARR2:LIST:METH RRC
:ACP:CARR2:LIST:METH?
Notes Maximum of Array length depends on the number of carriers
Couplings This command is an alias of:
[:SENSe]:ACPower:CARRier[1]|2:LIST:FILTer[:RRC][:STATe]
The enum value translates as follows:
–
RRC Weighted = 1|ON
–
Integ BW = 0|OFF
Maximum of Array length depends on the number of carriers
Preset
Modes Value
SA
IBW
WCDMA
RRC
LTEAFDD, LTEATDD, MSR
IBW
State Saved Saved in instrument state
Backwards
Compatibility
SCPI
[:SENSe]:ACPower:CARRier[1]|2:LIST:METHod IBW | RRC, …
[:SENSe]:ACPower:CARRier[1]|2:LIST:METHod?
Spectrum Analyzer Mode User's &Programmer's Reference 1095