3 Spectrum Analyzer Mode
3.12 List Sweep Measurement (Remote Only)
Preset Current electronic attenuation value from Swept SA measurement
State Saved Saved in instrument state
Range Same as the electronic attenuator range.
Remote
Command
[:SENSe]:LIST:EATTenuation:POINts?
Example
LIST:EATT:POIN?
3.12.3.4 RBW Type List
This command defines a list of analyzer RBW Type settings used for the
measurements.
The query form queries the analyzer for the values in the RBW Type list.
Remote
Command
[:SENSe]:LIST:BANDwidth|BWIDth:RESolution:TYPE <type>[, <type>]
[:SENSe]:LIST:BANDwidth|BWIDth:RESolution:TYPE?
Example
LIST:BAND:RES:TYPE GAUS
For this example, Types from Swept SA measurement include: GAUSsian|FLATtop|EMI
Preset Current RBW Type setting from Swept SA measurement
State Saved Saved in instrument state
Range Same as the RBW Types available in Swept SA measurement.
Remote
Command
[:SENSe]:LIST:BANDwidth|BWIDth:RESolution:TYPE:POINts?
Example
LIST:BAND:RES:TYPE:POIN?
3.12.3.5 RBW List
This command defines a list of analyzer RBW settings at which the measurements
are made.
The query form queries the analyzer for the values in the RBW list.
Remote
Command
[:SENSe]:LIST:BANDwidth|BWIDth:RESolution <freq>[, <freq>]
[:SENSe]:LIST:BANDwidth|BWIDth:RESolution?
Example
LIST:BAND:RES 100KHZ
Preset Current RBW value from Swept SA measurement
1938 Spectrum Analyzer Mode User's &Programmer's Reference