3 Spectrum Analyzer Mode
3.12 List Sweep Measurement (Remote Only)
Remote
Command
[:SENSe]:LIST:FREQuency:POINts?
Example
LIST:FREQ:POIN?
3.12.3.2 Mechanical Attenuator List
This command defines a list of analyzer mechanical attenuator settings at which the
measurements are made.
NOTE
Changing the mechanical attenuator within the list should be avoided if possible,
because it slows down the measurement and wears out the attenuator.
The query form queries the analyzer for the values in the mechanical attenuation
list.
Remote Command
[:SENSe]:LIST:ATTenuation <power>[, <power>]
[:SENSe]:LIST:ATTenuation?
Example
LIST:ATT 10DB
Dependencies If the requested setting is not valid, it is ignored and the previous setting is retained.
Preset Current mechanical attenuation value from Swept SA measurement
State Saved Saved in State
Range Same as the mechanical attenuator range.
Remote
Command
[:SENSe]:LIST:ATTenuation:POINts?
Example
LIST:ATT:POIN?
3.12.3.3 Electronic Attenuation List
This command defines a list of analyzer electronic attenuator settings at which the
measurements are made.
The query form queries the analyzer for the values in the electronic attenuation list.
Remote Command
[:SENSe]:LIST:EATTenuation <power>[, <power>]
[:SENSe]:LIST:EATTenuation?
Example
LIST:EATT 10DB
Dependencies If the required hardware option is not present, an error message would be issued. However, the error
message does not prevent list sweep from execution if all other list settings are set properly.