6 Input/Output
6.1 RF Source
Step Configuration of Step Duration parameter list (Remote Command Only)
Configures the Step Duration parameter array of the whole List Sequencer at one
time. The number of arrays is same as step number defined in "Number of Steps" on
page 2239. As a step is set up, the value entered runs through several levels of
validation.
Remote Command
:SOURce:LIST:SETup:DURation:TYPE <enum>, <enum>, <enum>,
:SOURce:LIST:SETup:DURation:TYPE?
Example
:SOUR:LIST:SET:DUR:TYPE COUN,TIME,CONT
:SOUR:LIST:SET:DUR:TYPE?
Notes Sets up or queries the parameter array of whole list sequence
Step Duration <enum> - specifies the duration of the step. The duration can be specified to be either
time, or play count of the ARB file associated with the step, or continuous. If Waveform is set to “CW”,
this value cannot be set to Play Count and an error will be generated. If continuous is selected, the
following Time or Count value is ignored. For further details of this setting, see "Step Duration" on
page 2249
Options are:
TIME | COUNt | CONTinuous
If input parameter number exceeds the step number defined by "Number of Steps" on page 2239,
then error -221, "Settings conflict; The number of input parameters is too large and is truncated to
current list step number" is generated, and only those parameters whose index number falls within
number of steps will be updated
Dependencies The range is 1 to 1000, which is determined by the number of steps you have configured. For details
see "Number of Steps" on page 2239
Step Configuration of Duration Time or Play Count parameter list (Remote Command
Only)
Configures the Duration Time or Play Count parameter array of the whole List
Sequencer at one time. The number of arrays is same as step number defined in
"Number of Steps" on page 2239. As a step is set up, the value entered runs through
several levels of validation.
Remote Command
:SOURce:LIST:SETup:TOCount <time/int>, <time/int>, <time/int>,
:SOURce:LIST:SETup:TOCount?
Example
:SOUR:LIST:SET:TOC 1s,2s,3s
:SOUR:LIST:SET:TOC?
:SOUR:LIST:SET:TOC 5,6,7
:SOUR:LIST:SET:TOC?
Notes Sets up or queries the parameter array of whole list sequence
Spectrum Analyzer Mode User's &Programmer's Reference 2259

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals