3 Spectrum Analyzer Mode
3.9 SEM Measurement
Preset Set to OFF by Global Settings > Restore Defaults and System > Restore Defaults > All Modes
Range
ON|OFF
Restore Defaults
Resets all functions in the Global settings menu to OFF. Pressing System, Restore
Defaults, All Modes has the same effect.
Remote Command
:INSTrument:COUPle:DEFault
Example
:INST:COUP:DEF
Backwards Compatibility
SCPI
:GLOBal:DEFault
3.9.19 Sweep
Accesses controls to configure and control the acquisition of data, and the X-axis
parameters of the instrument.
Depending on the selected mode and measurement, these controls might include:
Sweep Time, Continuous/Single, Pause/Resume, X Scale and Number of Points.
3.9.19.1 Sweep/Control
Accesses controls that let you operate the sweep and control functions of the
instrument, such as Sweep Time and Continuous/Single.
Sweep/Measure
Lets you toggle between Continuous and Single sweep or measurement operation.
The single/continuous state is Meas Global, so the setting affects all measurements.
The front-panel key Single/Cont performs exactly the same function
See "More Information" on page 1726
Remote
Command
:INITiate:CONTinuous OFF | ON | 0 | 1
:INITiate:CONTinuous?
Example Put instrument into Single measurement operation:
:INIT:CONT 0
:INIT:CONT OFF
Put instrument into Continuous measurement operation:
:INIT:CONT 1
Spectrum Analyzer Mode User's &Programmer's Reference 1725