4 System
4.6 Alignments
As with the PC6 and PC7, at every power-on, the software compares the timestamp
of the backup on the flash and the timestamp of the alignment data in use on the
disk. If the backup on the flash has newer data, you are given the opportunity to
restore the data from the backup before the application is loaded. If you respond
Yes, the Backup and Restore Alignment Data wizard (see "Alignment Data Wizard
(with Flash)" on page 2164) will be invoked and will prompt you to restore that
backup.
For purposes of these instructions, “alignment data” and “calibration data” are used
interchangeably.
Dependencies Not available in UXM
Remote
Command
:CALibration:DATA:DEFault
Example
:CAL:DATA:DEF
Notes Restores the alignment data files to their default state
Couplings Sets Auto Align to OFF. Sets Bit 14 in the Status Questionable Calibration register. The Error Condition
message “Align Now All required” is generated
Alignment Data Wizard (without Flash)
Guides you through the operation of backing-up or restoring the alignment data.
Spectrum Analyzer Mode User's &Programmer's Reference 2154