EasyManuals Logo

Keysight x-series User Manual

Keysight x-series
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #2808 background imageLoading...
Page #2808 background image
9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Remote Command
:STATus:QUEStionable:CALibration:CONDition?
Example
:STAT:QUES:CAL:COND?
Preset 0
Status Bits/OPC depend-
encies
Sequential command
Questionable Calibration Enable
Determines which bits in the "Questionable Calibration Condition" on page 2807
Register will set bits in the "Questionable Calibration Event Query" on page 2808
register, which also sets the Calibration Summary bit (bit 8) in the "Questionable
Register" on page 2794.
The variable <integer> is the sum of the decimal values of the bits you want to
enable.
Remote
Command
:STATus:QUEStionable:CALibration:ENABle <integer>
:STATus:QUEStionable:CALibration:ENABle?
Example
:STAT:QUES:CAL:ENAB 16384
Can be used to query if an alignment is needed, if you have turned off the automatic alignment process
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Questionable Calibration Event Query
Returns the decimal value of the sum of the bits in the Questionable Calibration
Event register.
NOTE
The register requires that the associated PTRansition or NTRansition filters be
set before a condition register bit can set a bit in the event register. The data in
this register is latched until it is queried. Once queried, the register is cleared.
Remote Command
:STATus:QUEStionable:CALibration[:EVENt]?
Example
:STAT:QUES:CAL?
Preset 0
Status Bits/OPC depend-
encies
Sequential command
Spectrum Analyzer Mode User's &Programmer's Reference 2808

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals