EasyManuals Logo

Keysight x-series User Manual

Keysight x-series
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1847 background imageLoading...
Page #1847 background image
3 Spectrum Analyzer Mode
3.10 TOI Measurement
When in Clear/Write, if a measurement-related instrument setting is changed when
the instrument is sweeping, a new sweep is initiated but the trace is not cleared.
While the sweep always uses the peak detector, in zero-span measurement the
average detector is used, and the metrics are calculated using the average of the
zero-span data.
Max Hold
The instrument maintains and displays a max hold trace, which represents the
maximum data value on a point-by-point basis of the new trace data and previous
trace data. Details of the count limiting behavior may be found under "Avg/Hold
Num" on page 1828 in Meas Setup.
Pressing Max Hold, or sending :TRAC:TYPE MAXH, sets the trace type to Max Hold,
causes the trace to be cleared, and causes the Max Hold sequence to be (re)started,
even if you are already in Max Hold. The max hold trace uses peak detection; in
zero-span measurement the metrics are calculated using the peak value of the
zero-span data.
Restarting the Max Hold sequence means:
The average/hold count k is set to 1, so that the next time the max hold trace is
displayed it simply represents one trace of new data
A new sweep is initiated
Clear and Write | Restart Averaging | Restart Max/Min Hold
Starts the trace writing, as though the "Trace Type" on page 1927 had just been
selected. The effect is exactly the same as reselecting the current Trace Type again
the control is provided because it may not be obvious that reselecting the same
selection from a radio button menu will take any action.
This control displays different labels, depending on the selected Trace Type:
Clear/Write: Clear and Write
Trace Average: Restart Averaging
Max Hold: Restart Max Hold
Min Hold: Restart Min Hold
Spectrum Analyzer Mode User's &Programmer's Reference 1847

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals