3 Spectrum Analyzer Mode
3.12 List Sweep Measurement (Remote Only)
Preset 0.0 s
State Saved Saved in State
Min 0 s
Max 100 s
Remote Command
[:SENSe]:LIST:TRIGger:HOLDoff:POINts?
Example
LIST:TRIG:HOLD:POIN?
Remote Command
Notes
Only one trigger holdoff selection is allowed.
3.12.3.13 Trigger Level
Sets the value at which the selected trigger input will trigger a new sweep.
Remote Command
[:SENSe]:LIST:TRIGger:LEVel <ampl>
[:SENSe]:LIST:TRIGger:LEVel?
Example
LIST:TRIG:LEV 0.4V
Preset 1.2 V
State Saved Saved in instrument state
Min -5 V
Max 5 V
Remote Command
[:SENSe]:LIST:TRIGger:LEVel:POINts?
Example
LIST:TRIG:LEV:POIN?
Remote Command
Notes
Only one trigger level selection is allowed.
3.12.3.14 Trigger Slope
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative
to trigger on a falling edge.
Remote Command
[:SENSe]:LIST:TRIGger:SLOPe POSitive | NEGative
[:SENSe]:LIST:TRIGger:SLOPe?
Example
LIST:TRIG:SLOP NEG
Preset POSitive
1944 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals