3 Spectrum Analyzer Mode
3.9 SEM Measurement
Avg/Hold Num
Toggles averaging On or Off, in addition to enabling you to set the number of
measurement averages used to calculate the measurement result. The average is
displayed at the end of each sweep. After the specified number of average counts,
the average mode (termination control) setting determines the average action.
In the remote mode, use "Averaging On/Off" on page 1621 to turn Averaging on or
off.
Remote Command
[:SENSe]:SEMask:AVERage:COUNt <integer>
[:SENSe]:SEMask:AVERage:COUNt?
Example
:SEM:AVER:COUN 100
:SEM:AVER:COUN?
Preset 10
State Saved Saved in instrument state
Min/Max 1/10000
Averaging On/Off
Turns Averaging on or off.
NOTE
In this measurement, Average Type is preset to the Log-Pwr Avg (Video)
method. Other averaging methods are not available.
Remote Command
[:SENSe]:SEMask:AVERage[:STATe] ON | OFF | 1 | 0
[:SENSe]:SEMask:AVERage[:STATe]?
Example
:SEM:AVER ON
:SEM:AVER?
Preset
OFF
State Saved Saved in instrument state
Range
ON|OFF
Meas Method
Sets the measurement method:
Method Option Description
Integration BW
0|OFF
Enables you to set the channel integration bandwidth
RRC Weighted
1|ON
Selects Root Raised Cosine (RRC) filtering of the carriers. The a
Spectrum Analyzer Mode User's &Programmer's Reference 1621