9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Remote Command
:STATus:OPERation:CONDition?
Example
:STAT:OPER:COND?
Preset 0
Status Bits/OPC depend-
encies
Sequential command
Operation Enable
Determines which bits in the "Operation Event Query" on page 2789 register will set
the Operation Status Summary bit (bit 7) in the "Status Byte Register" on page 2781.
The variable <integer> is the sum of the decimal values of the bits you want to
enable.
NOTE
The preset condition is to have all bits in this enable register set to 0. To have any
Operation Events reported to the Status Byte Register, one or more bits need to
be set to 1.
Remote
Command
:STATus:OPERation:ENABle <integer>
:STATus:OPERation:ENABle?
Example
:STAT:OPER:ENAB 1
Sets the register so that Align Now events will be reported to the Status Byte Register
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Operation Event Query
Returns the decimal value of the sum of the bits in the Operation Event register.
NOTE
The register requires that the associated PTRansition or NTRansition filters be
set before a condition register bit can set a bit in the event register. The data in
this register is latched until it is queried. Once queried, the register is cleared.
Remote Command
:STATus:OPERation[:EVENt]?
Example
:STAT:OPER?
Preset 0
2789 Spectrum Analyzer Mode User's &Programmer's Reference