3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
Image Protection
Lets you turn IF Protection on and off. IF Protection is a technique used in
unpreselected instruments (such as Keysight’s modular instruments) to detect and
suppress images and spurs that may be present in non-preselected hardware.
IF Protection takes two sweeps and by correlating the data between them, provides
a single, correct power-versus-frequency trace.
Remote Command
[:SENSe]:SWEep:IMAGeprot OFF | ON | 0 | 1
[:SENSe]:SWEep:IMAGeprot?
Dependencies Only appears in SA Mode in VXT Model M94. Does not appear in VXT Models M9410A/11A
Grayed-out in Zero Span
Preset
ON
State Saved Saved in instrument state
504 Spectrum Analyzer Mode User's &Programmer's Reference