EasyManuals Logo

Keysight x-series User Manual

Keysight x-series
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1057 background imageLoading...
Page #1057 background image
3 Spectrum Analyzer Mode
3.5 ACP Measurement
Type Option Description
evaluated
Max Power
Carrier
MPCarrier
Maximum carrier power among the carriers of Measure Carrier On is the
reference of measurement. If Measure Carriers of all the carriers are off, the
reference power and all the relative power results are NAN. Relative limits are
not evaluated
Min Power
Carrier
5G NR only
MINPcarrier
Minimum carrier power among the carriers of Measure Carrier On is the
reference of measurement. If Measure Carriers of all the carriers are off, the
reference power and all the relative power results are NAN. Relative limits are
not evaluated
Carrier Index
CINDex
Power of the specified carrier is the reference of measurement. If Measure
Carriers of this carrier index is off, the reference power and all the relative
power results are NAN. Relative limits are not evaluated
Manual
MANual
Power or PSD specified by the user is the reference of measurement
Aggregated
Chan BW
LTEAFDD,
LTEATDD, 5G NR
only
ACBandwidth
The assigned aggregated channel bandwidth power which is measured with a
rectangular filter with measurement bandwidth specified as aggregated
channel bandwidth based on the definition of each 3GPP standard. Calculated
from the carrier configuration including SCS (Power Meas), the smallest SCS
among the enabled SCSs of the selected component carrier. If Measure Carriers
of all the carriers are off, the reference power and all the relative power results
are NAN. Relative limits are not evaluated
Max Power
Carrier in Sub-
block
MPCSubblock
Maximum carrier power among the sub-block carriers of Measure Carrier On is
the reference of measurement. If Measure Carriers of all the carriers in a sub-
block are off, the reference power of the sub-block and all the relative power
results referring to this sub-block are NAN, and these relative limits are not
evaluated
Total
Multicarriers
MSR only
TMCarriers
Total power of multi carriers is the power reference of measurement. Each
carrier power is calculated with its own carrier configuration settings
Min Power
Carrier in Sub-
block
5G NR only
MINSubbloc
Minimum carrier power among the sub-block carriers of Measure Carrier On is
the reference of measurement. If Measure Carriers of all the carriers in a sub-
block are off, the reference power of the sub-block and all the relative power
results referring to this sub-block are NAN, and these relative limits are not
evaluated
Left & Right Sub-
blocks
5G NR only
LRSubblocks
The reference depends on the number of Component Carriers (CC) and Carrier
Allocation as follows:
Num of CC is 1: the carrier power is the reference
Num of CC is 2 or more & Carrier Allocation is Contiguous: Aggregated
Channel power is the reference
Num of CC is 2 or more & Carrier Allocation is Non-Contiguous: Aggregated
powers of left and right sub-blocks are the references. Left and right sub-
blocks are determined by component carrier configuration
Spectrum Analyzer Mode User's &Programmer's Reference 1057

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)