EasyManuals Logo

Keysight x-series User Manual

Keysight x-series
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1121 background imageLoading...
Page #1121 background image
3 Spectrum Analyzer Mode
3.5 ACP Measurement
The values of the trace points are assumed to be in a decibel scale, as they are
internally stored.
If a point in FirstTrace is equal to maxtracevalue, the resultant point is also
maxtracevalue.
Otherwise, if the result of the subtraction is less than or equal to 0, the resultant
point is mintracevalue.
Power Sum (Op1 + Op2)
Calculates a power sum between the First Trace operand and the Second
Trace operand and puts the result in the destination trace.
During the sweep, the following formula is executed for each point in the trace
operands, and the corresponding point is generated for the destination trace.
DestinationTrace = 10 log(10(1/10)(FirstTrace) + 10(1/10)(SecondTrace))
The values of the trace points are assumed to be in a decibel scale, as they are
internally stored.
If a point in either trace operand is equal to maxtracevalue, the resultant point is
also maxtracevalue.
Log Offset (Op1 + Offset)
Calculates a log offset from the First Trace operand and puts the result in the
destination trace. This is like the B-DL function in some older instruments. The offset
is entered on the Offset control, which only appears when this math function is in
force for the selected trace. Each destination trace has its own offset.
During the sweep, the following formula is executed for each point in the trace
operand, and the corresponding point is generated for the destination trace.
DestinationTrace = FirstTrace + Offset
The values of the trace points are assumed to be in dBm (as they are internally
stored) and the offset is in dB.
If a point in the trace operand is equal to maxtracevalue, the resultant point is also
maxtracevalue.
If a point in the trace operand is equal to mintracevalue, the resultant point is also
mintracevalue.
Example: If offset is 25 dB, then our destination trace will be higher than the
operand trace by 25 dB.
Note that the Second Trace operand is not used for this function.
Spectrum Analyzer Mode User's &Programmer's Reference 1121

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals