3 Spectrum Analyzer Mode
3.7 Burst Power Measurement
frequency is reflected in the disabled message displayed for Electrical
Attenuator. For N9032B and N9042B IFBW values > 1.5GHz do not support a
Center Frequency that can reach the Low Band maximum frequency. In these
cases, the Electrical Attenuator will remain disabled no matter the Center
Frequency.
More Information
The electronic attenuator offers finer steps than the mechanical attenuator, has no
acoustical noise, is faster, and is less subject to wear. These advantages primarily
aid in remote operation and are negligible for front panel use. See "Using the
Electronic Attenuator: Pros and Cons" on page 1259 for a detailed discussion of the
pros and cons of using the electronic attenuator.
For the Single-Attenuator configuration, for SCPI backwards compatibility, the
“soft” attenuation feature replaces the Dual-Attenuator configuration’s electronic
attenuator. All the same couplings and limitations apply. See "Attenuator
Configurations and Auto/Man" on page 1959
Mechanical Attenuator Transition Rules
When the Electronic Attenuator is enabled, the Mechanical Attenuator transitions to
a state that has no Auto function. Below are the rules for transitioning the
Mechanical Attenuator. Note that the information below only applies to the Dual-
Attenuator configurations, and only when the Electronic Attenuator is installed:
When the Electronic Attenuation is enabled from a disabled state:
The Mechanical Attenuator is initialized to 10dB (this is its optimal performance
setting). You can then set it as desired with SCPI, numeric keypad, step keys, or
knob, and it behaves as it normally would in manual mode
The Auto/Man state of (Mech) Atten is saved
The Auto/Man toggle on the (Mech) Atten control disappears, and the auto rules
are disabled
The Electronic Attenuator is set to 10dB less than the previous value of the
Mechanical Attenuator, within the limitation that it must stay within the range of
0 to 24dB of attenuation
Examples in the Dual-Attenuator configuration:
Mech Atten at 20dB. Elec Atten enabled, Mech Atten set to 10dB, and Elec
Atten set to 10dB. New total attenuation equals the value before Elec Atten
enabled
1258 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?