EasyManuals Logo

Keysight x-series User Manual

Keysight x-series
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1353 background imageLoading...
Page #1353 background image
3 Spectrum Analyzer Mode
3.8 Spurious Emissions Measurement
n Return Value
1 or omitted Returns a variable-length (1 + 6 * Spurs up to 1201 entries) comma-separated list containing detailed
information in the following format:
Number of spurs in following list (Integer)
[Repeat the following for each spur]
Spur #
Range # Spur was located (Integer)
Frequency of Spur (Hz, Float64)
Amplitude of Spur (dBm, Float32)
Absolute Limit (dBm, Float32)
Pass or Fail (1|0, Boolean)
2 21 (Average
Trace)
Regardless of the Trace selection, returns a comma-separated list of the average trace data for the
selected range (where range number = n 1) using Detector 1. If selected range is not active, SCPI_
NAN is returned for each trace data element
22 Returns the number of spurs found for the selected Measured Trace
23 42 (Average
Trace)
Regardless of the Trace selection, returns a comma-separated list of the average trace data for the
selected range (where range number = n 22) using Detector 2. If selected range is not active or
Detector 2 selection is off, SCPI_NAN is returned for each trace data element
43 62
(Maximum Hold
Trace)
Regardless of the Trace selection, returns a comma-separated list of the maximum hold trace data for
the selected range (where range number = n 42) using Detector 1. If selected range is not active,
SCPI_NAN is returned for each trace data element
63 82
(Minimum Hold
Trace)
Regardless of the Trace selection, returns a comma-separated list of the minimum hold trace data for
the selected range (where range number = n 62) using Detector 1. If selected range is not active,
SCPI_NAN is returned for each trace data element
83-102 Reserved
103-122 Reserved
123-142
(Clear/Write
Trace)
Regardless of the Trace selection, returns a comma-separated list of the clear/write trace data for the
selected range (where range number = n 122) using Detector 1. If selected range is not active,
SCPI_NAN is returned for each trace data element
143-162
(Clear/Write
Trace)
Regardless of the Trace selection, returns a comma-separated list of the clear/write trace data for the
selected range (where range number = n 142) using Detector 2. If selected range is not active or
Detector 2 selection is off, SCPI_NAN is returned for each trace data element
163-182 (Trace
1)
Returns a comma-separated list of the trace data of Trace 1 for the selected range (where range
number = n 162). If selected range is not active, SCPI_NAN is returned for each trace data element
183-202 (Trace
2)
Returns a comma-separated list of the trace data of Trace 2 for the selected range (where range
number = n 182). If selected range is not active, SCPI_NAN is returned for each trace data element
203-222 (Trace
3)
Returns a comma-separated list of the trace data of Trace 3 for the selected range (where range
number = n 202). If selected range is not active, SCPI_NAN is returned for each trace data element
223 Returns "Marker Table" on page 1358 data as a series of comma separated values in the following form:
Spectrum Analyzer Mode User's &Programmer's Reference 1353

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals