6 Input/Output
6.5 Corrections
Remote Command
[:SENSe]:CORRection:CSET:GROup:BReak ON | OFF | 1 | 0
[:SENSe]:CORRection:CSET:GROup:BReak?
Example
:SENS:CORR:CSET:GRO:BR ON
Notes When running the Frequency Scan measurement in EMI Receiver Mode, if break is turned on when a
SCPI is sent to start the scan, the scan pauses when it reaches the boundary of correction group
ranges. Bit 8 (Paused) of status operation register is set to true. To resume, send :INITi-
ate2:RESume. To stop the scan, send :ABORt
When running the Swept SA measurement in Spectrum Analyzer Mode, the break state does not
affect the operation of sweep when SCPI to control the sweep is sent. Instead, the SCPI commands
close the message prompt if it is showing at the point the commands are sent, and the break is turned
off. The SCPI includes:
:INITitate:IMMEdiate
:INITitate:RESTart
:INITitate:CONTinuous ON | OFF | 1 | 0
:ABORt
Dependencies Correction group is supported in EMI Receiver Mode, and in Spectrum Analyzer Mode if option EMC
or EMI Receiver mode is present. If you switch to other measurements or modes, correction group is
turned off and the Correction Group functions (like Break) are not visible
Preset Not affected by Preset. Set to OFF by Restore Input/Output Defaults
State Saved Saved in instrument state
Spectrum Analyzer Mode User's &Programmer's Reference 2441