7 Save/Recall/Print
7.3 Save
Similarly, the Bias rows can only be used in Correction register 1, because there can
only be one setting for Bias at any given time. If a Correction file with a Bias or Bias
State row is loaded into any Correction register but 1, an error is generated: Mass
storage error; Can only load Bias Settings into Correction 1
The data follows the DATA row, as comma separated X, Y pairs; one pair per line.
For example, suppose you have an Antenna to correct for on an N9020A version
A.02.06 and the correction data is:
–
0dB at 200MHz
–
17dB at 210MHz
–
14.8dB at 225MHz
Then the file will look like:
–
Amplitude Correction
–
"Correction Factors for 11966E"
–
"Class B Radiated"
–
A.02.06,N9020A
–
P13 EA3 UK6,01
–
Frequency Unit, MHz
–
Antenna Unit, dBuV/m
–
Frequency Interpolation, Linear
–
DATA
–
200.000000,0.00
–
210.000000,17.00
–
225.000000,14.80
The choices for the 1 of N fields in the metadata are as follows:
–
Frequency Unit: Hz, kHz, MHz, GHz
–
Antenna Unit: dBuV/m, dBuA/m, dBG, dBpT, None
–
Frequency Interpolation: Logarithmic, Linear