8 Trigger
8.6 Auto/Holdoff
having been below the threshold for at least the holdoff time. For negative slope, the
trigger event is generated if the signal characteristic crosses the threshold (with
negative slope) and then remains below the threshold for at least the holdoff time. In
either case, the trigger event is associated with the time the level was crossed
Remote Command
:TRIGger[:SEQuence]:HOLDoff:TYPE NORMal | ABOVe | BELow
:TRIGger[:SEQuence]:HOLDoff:TYPE?
Example
:TRIG:HOLD:TYPE NORM
Preset
Modes Setting
GSM/EDGE
Bluetooth
BELow
All others
NORMal
State Saved Saved in instrument state
2703 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals