3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
–
Normal, Delta, Fixed, and Off work as for the standard menu above
–
Marker Properties opens the Marker Properties tab (see "Properties" on page
335). Move Marker -> Display Trace and Move Display Trace -> Marker work the
same way as the corresponding buttons in the Marker -> menu (see "Marker To"
on page 370)
–
Peak Search finds the highest peak on the current Display Trace
–
Peak Search all Traces finds the highest peak in the Waterfall window (see "Peak
Search All Traces" on page 321)
NOTE
In earlier HP/Agilent/Keysight analyzers, markers stayed at the same position on
the display even when you changed frequency. In the X-Series, markers stay at
the frequency they are set to, even if you change Center Frequency. So, your
marker will move, possibly offscreen, when you change frequency. This is a
308 Spectrum Analyzer Mode User's &Programmer's Reference