3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
placed
The response to the query is 0 if OFF, 1 if ON
Couplings The state of Auto Initialize is not affected by "Auto Couple" on page 2017
Auto Initialize is set to ON by Preset or All Markers Off
If Auto Initialize is ON for a marker and that marker is on, that marker’s Marker Trace is immediately set
according to the above flowchart
Sending the remote command causes the addressed marker to become selected
Preset
ON
When the marker moves between traces the marker’s X position in trace points is
retained as it moves. For moving between active traces this generally means the x-
axis value of the marker does not change. But for moving to or from an inactive
trace, the x-axis value takes on that of the new trace at the bucket the marker was
on the old trace (and is still on, on the new trace, since the bucket doesn’t change).
Note this is true even if the marker is off screen. Thus, a marker that is at the center
of the screen on the old trace stays at the center of the screen on the new trace. A
marker that is off screen one whole screen to the left on the old trace remains off
screen one whole screen to the left on the new trace even if this means it is at
negative time!
Auto Init Rules Flowchart
The following flowchart depicts the Auto Initialize rules:
340 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals