3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
Mixing is ON. Since all this happens between sweeps, you never see the instrument
retuning to do the counts.
If you wish to see the entered frequency of a counted marker it will appear in the
active function area when that marker is selected (for Fixed markers, you must press
Marker, Fixed to select Fixed markers, then press it a second time to view or adjust
the x or y marker values).
Counting Off-screen Markers
If the selected marker is Off, the X-axis the instrument can still be tuned to the
marker (unless it is outside the current range of the instrument), so the count can
still be displayed. This means you can see a count for an off-screen marker, even
though there may be no valid Y-value for the marker. If the marker frequency is
outside the range of the instrument, the display shows three dashes in the count
block (---), and Not a Number (NAN) is returned to a SCPI count query.
Delta Marker
When a Delta Marker is selected while Marker Count is ON:
If the reference marker is not a Fixed marker, the display shows the difference
between the count of the selected marker and the count of the reference marker
If the reference marker is a Fixed marker and there is a count stored in the
marker (because Marker Count was ON when the marker became a Fixed
marker), the display shows the difference between the count at the marker and
the count stored in the reference marker
Marker Count works in Zero Span as well as in Swept SA. The instrument tunes to
the frequency of the selected marker, which, for active zero span traces, is simply
the center frequency of the instrument.
Fixed Markers
Fixed markers have a count stored in them that is generally kept fixed and not
updated. If a Fixed marker is selected, or used as a reference, the signal at the
marker frequency is not counted; rather the stored count is seen or used as the
reference. The count is stored, if Count is ON, when the marker becomes Fixed or
when, while Fixed, the marker is moved by re-zeroing the reference (if it is the
reference marker) or via a peak search (since both, by definition, use valid trace
data). The count stored in a Fixed marker is lost if the counter is turned off, if the
marker is moved to an inactive trace, or if the marker is moved by adjusting its x-
value.
378 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals