3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
instruments are capable of making Quasi Peak and EMI Average detected
measurements correctly on a log scale, due to the digital IF. This latter capability
means that the user can observe detected EMI levels on a log scale, allowing a large
visible dynamic range. You can manually set the Average Type to Log-Power or
Power, but the results will no longer be CISPR compliant.
Also, in the past, EMI analysis equipment would need to perform a ranging
operation to set the reference level when one of these detectors was turned on, but
the X-series instruments do not - because of its digital IF, there is no need to set the
reference level (range) to improve the accuracy nor to allow visibility of the detected
level.
EMI Average
The EMI Average detector in Agilent’s X-Series instruments is so called to
distinguish it from the Average detector, although EMI users typically refer to it
simply as the “Average detector”. The intent of this detector is to provide a standard
means to “smooth” the signal while still providing compliance to CISPR pulse
response standards.
Unlike the regular Average detector, which averages on a bucket-by-bucket basis
using either a power, log-power or voltage scale (a bucket is the same as a trace
point), the EMI Average detector displays the average value, on the voltage scale, of
the overall amplitude envelope, independent of the trace bucket width. It is defined
for EMI measurements by the CISPR 16–1–1 standard and, in the X-series, uses a
sophisticated algorithm to implement a lowpass filter that conforms to the latest
CISPR standard.
Note that CISPR standard operation is to perform the envelope averaging on the
voltage scale. You can manually set the Average Type to Log-Power or Power, but
the results will no longer be CISPR compliant. See note under Quasi-Peak.
RMS Average
Not to be confused with the RMS mode of the regular Average detector, this is a
special filter for making EMI measurements. It is a frequency dependent
RMS/Averaging filter, used in making CISPR compliant EMI measurements. This
filter conforms to the 2007 revision of the CISPR 16–1–1 standard. This detector
does one averaging process (in the VBW hardware) on the "power" (a.k.a. RMS)
scale and another process on the voltage scale using a "meter movement simulator"
similar to the one used in the QPD filter.
Note that CISPR standard operation is to perform the envelope averaging on the
voltage scale. You can manually set the Average Type to Log-Power or Power, but
the results will no longer be CISPR compliant. See note under Quasi-Peak.
Spectrum Analyzer Mode User's &Programmer's Reference 527