3 Spectrum Analyzer Mode
3.4 Occupied BW Measurement
Example Set Center Frequency to 50MHz:
:FREQ:CENT 50 MHz
Increment Center Frequency by the value of CF Step:
:FREQ:CENT UP
Return the current value of Center Frequency:
:FREQ:CENT?
Notes Sets the RF, External Mixing or I/Q Center Frequency depending on the selected input:
For RF input it is equivalent to :FREQ:RF:CENT
For I/Q input it is equivalent to :FREQ:IQ:CENT
For External Mixer it is equivalent to :FREQ:EMIX:CENT
Preset and Max values depend on Hardware Options
If no terminator (for example, MHz) is sent, the terminator Hz is used. If a terminator with unit other
than frequency is used, an invalid suffix error message is generated
Dependencies Center Frequency can be limited by Start Freq or Stop Freq limits, if Span is so large that start or stop
reach their limits
Couplings When operating in “swept span”, any value of Center Frequency or Span that is within the frequency
range of the instrument is allowed, if the value is being set through the front panel numeric key pad or
the SCPI command. The other parameter is forced to a different value if needed, to keep Start and
Stop Frequencies within the instrument’s frequency range
Coupling between center frequency and span: numeric (keypad) entries are treated differently than
changing the value using the step keys (Up/Down Arrows) or the knob. Similarly, for remote operation,
sending a numeric frequency value is treated differently than the UP|DOWN keywords:
Numeric entries (keypad or remote): Any value of Center Frequency or Span (within the frequency
range of the instrument) is allowed. The other parameter is changed, as necessary, to keep the
Start Freq and Stop Freq within the instrument frequency range
Knob or Step keys (up/down arrows) or UP|DOWN keywords: The value of the parameter being
changed (Center Frequency or Span) is limited so the other parameter is not forced to a new
value. Thus, if only the step keys and knob are used, you can return to the initial Center Frequency
and Span by changing only the current parameter
Note that, since out-of-range Start Freq and Stop Freq are never allowed, markers and trace math
work correctly without requiring any special handling for out-of-range conditions
Preset Depends on instrument maximum frequency, mode, measurement, and selected input
See "Center Frequency Presets" on page 785 and "Ext Mix Center Freq" on page 787
State Saved Saved in instrument state
Min/Max Depends on instrument maximum frequency, mode, measurement, and selected input
See "Center Frequency Presets" on page 785 and "Ext Mix Center Freq" on page 787
Annotation Center <value> appears in the lower left corner of the display
Status Bits/OPC
dependencies
Non-overlapped
784 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals