EasyManuals Logo

Keysight x-series User Manual

Keysight x-series
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #823 background imageLoading...
Page #823 background image
3 Spectrum Analyzer Mode
3.4 Occupied BW Measurement
As shown in the table above, the On state (in Modes that do not support Adaptive
NFE) matches the Full state in Modes that do support Adaptive NFE.
To maintain SCPI backwards compatibility, the existing command to turn NFE on or
off is retained, and a new command is added to set the state to turn AdaptiveON or
OFF:
[:SENSe]:CORRection:NOISe:FLOor ON|OFF|1|0 is retained, with the
default changed to ON for Modes that support Adaptive NFE
[:SENSe]:CORRection:NOISe:FLOor:ADAPtive ON|OFF|1|0 is added (for
certain Modes), default = ON
When NFE is On or Full, the expected noise power of the instrument (derived from a
factory calibration) is subtracted from the trace data. This will usually reduce the
apparent noise level by about 10dB in low band, and 8dB in high band (>~3.6GHz).
NFE works with any RBW, VBW, detector, any setting of Average Type, any amount
of trace averaging, and any signal type. It is ineffective when the trace is not
smoothed (smoothing processes include narrow VBWs, trace averaging, and long
sweep times with the detector set to Average or Peak). It works best with extreme
amounts of smoothing, and with the average detector, with the Average Type set to
Power.
In those cases where the cancellation is ineffective, it nonetheless has no
undesirable side-effects. There is no significant speed impact to having Noise Floor
Extension on.
The best accuracy is achieved when substantial smoothing occurs in each point
before trace averaging. Thus, when using the Average detector, results are better
with long sweep times and fewer trace averages. When using the Sample detector,
the VBW filter should be set narrow with less trace averaging, instead of a wide
VBW filter with more trace averaging.
NOTE
Noise Floor Extension has no effect unless the RF Input is selected, so when
External Mixing is selected, it does nothing.
For more details, see "Optimal Detector & Averaging Selections" on page 825 and
"Recalibration of Noise Floor" on page 826.
Pros & Cons of Adaptive NFE
Adaptive NFE provides an alternative to fully-on or fully-off NFE. Fully-on NFE can,
notably in cases with little or no averaging of the spectrum, result in a display that is
distractingly unfamiliar in the variability in response to low level signals. Fully-off
NFE fails to achieve the potential improvement in dynamic range and associated
accuracy of measurement of low-level signals. Adaptive NFE controls the degree of
potential improvement in the noise floor to give more improvement for those
instrument settings that can make good use of the potential improvement—those
Spectrum Analyzer Mode User's &Programmer's Reference 823

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)