3 Spectrum Analyzer Mode
3.7 Burst Power Measurement
Remote Interface Examples
Command examples and details appear in the table below. Note that each of the
commands below sets the amplitude unit only for the selected amplitude scale (Log
or Lin), the other scale is unaffected.
Unit Example Notes
dBm :UNIT:POW
DBM
dB relative to one milliwatt
dBmV :UNIT:POW
DBMV
dB relative to one millivolt
dBmA :UNIT:POW
DBMA
dB relative to one milliamp
W :UNIT:POW W Watts
V :UNIT:POW V Volts
A :UNIT:POW A Amperes
dBmV :UNIT:POW
DBUV
dB relative to one microvolt
dBmA :UNIT:POW
DBUA
dB relative to one microamp
The unit dBuA can also appear as a "Transducer Unit" on page 1247
When querying the Y-Axis unit, you can query the Transducer Unit to distinguish
between regular dBuA and the dBuA transducer unit. If :CORR:CSET:ANT?
returns NOC (for No Conversion), you are using a normal Y Axis dBuA. If it returns
UA, you are using a Transducer Unit dBuA
dBpW :UNIT:POW
DBPW
dB relative to one picowatt
dBmV/m
(Transducer
Unit)
:UNIT:POW
DBUVM
Sets the amplitude unit for the selected amplitude scale (log/lin) to dBmV/meter.
This selection is only available if a Correction is turned on, and the Transducer Unit
for that Correction is not None
See "Transducer Unit" on page 1247
dBmA/m
(Transducer
Unit)
:UNIT:POW
DBUAM
Sets the amplitude unit for the selected amplitude scale (log/lin) to dBmA/meter.
This selection is only available if a Correction is turned on, and the Transducer Unit
for that Correction is not None
See "Transducer Unit" on page 1247
dBpT
(Transducer
Unit)
:UNIT:POW
DBPT
Sets the amplitude unit for the selected amplitude scale (log/lin) to dBpT (dB
relative to one picotesla). This selection is only available if a Correction is turned on,
and the Transducer Unit for that Correction is not None
See "Transducer Unit" on page 1247
dBG
(Transducer
Unit)
:UNIT:POW
DBG
Sets the amplitude unit for the selected amplitude scale (log/lin) to dBG (dB relative
to one Gauss). This selection is only available if a Correction is turned on, and the
Transducer Unit for that Correction is not None
See "Transducer Unit" on page 1247
dBmA :UNIT:POW Sets the amplitude unit for the selected amplitude scale (log/lin) to dBmA. This
Spectrum Analyzer Mode User's &Programmer's Reference 1245

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals