3 Spectrum Analyzer Mode
3.7 Burst Power Measurement
Unit Example Notes
(Transducer
Unit)
DBUA selection is only available as a Transducer Unit if a Correction is turned on, and the
Transducer Unit for that Correction is not None
See "Transducer Unit" on page 1247
The unit dBuA can also appear as a normal Y Axis Unit (see above)
dBuA as a Transducer Unit is used when using current probes, because current
probes are often supplied with conversion tables that provide the transducer
factors. When dBuA is used as a Transducer Unit, the normal conversion from
power to amps for dBuA (based on the instrument input impedance) is not done,
but instead the conversion is based solely on the Correction that contains the
transducer factors. This is what distinguishes dBuA as a normal unit from dBuA as a
transducer unit
When querying the Y-Axis unit, you can query the Transducer Unit to distinguish
between regular dBuA and the dBuA Transducer Unit. If :CORR:CSET:ANT?
returns NOC (for No Conversion), you are using a normal Y-Axis dBuA. If it returns
UA, you are using a Transducer Unit dBuA
Amplitude Data Query and Y Axis Unit
The settings of Y-Axis Unit and Display Scale affect how the data is returned over
the remote interface in response to a query. When using the remote interface, no
unit is returned, so you must know what the Y-Axis unit is to interpret the results:
Example 1
Set the following:
Display Scale (Log)
Y Axis Unit, dBm
Scale/Div, 1dB
Ref Level, 10dBm
This sets the top line to 10dBm with each vertical division representing 1dB. Thus, if
a point on trace 1 is on the fifth graticule line from the top, it represents 5dBm and
will read out remotely as 5.
Example 2
Set the following:
Display Scale (Lin)
Y Axis Unit, Volts
Ref Level, 100mV (10mV/div)
1246 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)