4 System
4.6 Alignments
“Time while Auto Align off” is not available in VXT models M9410A/11A.
A successful Align Now, RF sets the Last Align RF temperature to the current
temperature, and resets the Last Align RF time. A successful Align Now All or Align
Now All but RF sets the Last Align Now All temperature to the current temperature,
and resets the Last Align Now All time. A successful Align Now All also resets the
Last Align RF items if the RF portion of the Align Now succeeded.
Example
:SYST:SHOW ALIGN
Notes The values displayed on the screen are only updated upon entry to the screen, and not updated while
the screen is being displayed
The following data-specific queries are available:
Query Time since Startup
Remote Command
:SYSTem:PON:TIME?
Example
:SYST:PON:TIME?
Notes Value is the time since the most recent start-up in seconds
State Saved No
Query Current Temperature
Remote
Command
:CALibration:TEMPerature:CURRent?
Example
:CAL:TEMP:CURR?
Notes Value is in degrees Centigrade
State Saved No
Spectrum Analyzer Mode User's &Programmer's Reference 2132

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals