3 Short-Range Comms & IoT Mode
3.8 SEM Measurement
whether or not the Low Noise Path Enable is selected in the user interface. The only
time the Low Noise Path is used is when Low Noise Path Enable is selected, the
sweep is completely in High Band (> 3.6 GHz) and no preamp is in use.
For measurements that use IQ acquisition, the low noise path is used when Center
Frequency is in High Band (> 3.6GHz) and no preamp is in use. In other words, the
rules above are modified to use only the center frequency to qualify which path to
switch in. This is not the case for FFTs in the Swept SA measurement; they use the
same rules as swept measurements.
Note that the Low Noise Path, while giving improved DANL, has the disadvantage of
decreased TOI performance and decreased gain compression performance relative
to the standard path.
Note also that the bypass switch is a mechanical switch and has finite life, so if the
Low Noise Path Enable is selected, it is possible to cause frequent cycling of this
switch by frequently changing instrument settings such that the above conditions
hold true only some of the time. A user making tests of this nature should consider
opting for the Standard Path, which will never throw the bypass switch, at the
expense of some degraded noise performance.
The low noise path is useful for situations where the signal level is so low that the
instrument performance is dominated by noise even with 0 dB attenuation, but still
high enough that the preamp option would have excessive third-order
intermodulation or compression. The preamp, if purchased and used, gives better
noise floor than does the “Low Noise Path.” However, its compression threshold and
third-order intercept are much poorer than that of the non-preamp path.
There are some applications, typically for signals around −30dBm, for which the
third-order dynamic range of the standard path is good enough, but the noise floor
is not low enough even with 0dB input attenuation. When the third-order dynamic
range of the preamp path is too little and the noise floor of the standard path is too
high, the Low Noise Path can provide the best dynamic range
The graph below illustrates the concept. It shows, in red, the performance of an
instrument at different attenuation settings, both with the preamp on and off, in a
measurement that is affected by both instrument noise and instrument TOI. The
green shows the best available dynamic range, offset by 0.5dB for clarity. The blue
shows how the best available dynamic range improves for moderate signal levels
with the low noise path switched in. In this illustration, the preamp improves the
noise floor by 15dB while degrading the third-order intercept by 30dB, and the low
noise path reduces loss by 8dB. The attenuator step size is 2dB.
1070 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals