EasyManuals Logo

Keysight x-series Instruction Manual

Keysight x-series
2530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1194 background imageLoading...
Page #1194 background image
3 Short-Range Comms & IoT Mode
3.8 SEM Measurement
Preset
ON
State Saved Saved in instrument state
Range
OFF|ON
Backwards
Compatibility
SCPI
:DISPlay:SEMask:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:COUPle
3.8.19.3 Sweep Config
Accesses controls that enable you to configure the Sweep and Control functions of
the instrument, such as Sweep Rules.
Points
Sets the number of points displayed in the traces. The current value of points is
displayed in the bottom-right corner of the display.
Remote Command
[:SENSe]:SEMask:SWEep:POINts <integer>
[:SENSe]:SEMask:SWEep:POINts?
Example
:SEM:SWE:POIN 4001
:SEM:SWE:POIN?
Preset 2001
State Saved Saved in instrument state
Min 201
Max 10001
Annotation On second line of annotations in bottom right corner
IF Dithering
Lets you turn IF Dithering on or off. This is a technique used in unpreselected
instruments (such as Keysight’s modular instruments) to enhance the rejection of
images and internally-generated spurious signals.
Remote Command
[:SENSe]:SWEep:IF:DITHer OFF | ON | 0 | 1
[:SENSe]:SWEep:IF:DITHer?
Dependencies Only appears in Spectrum Analyzer Mode in VXT models
Preset
OFF
State Saved Saved in instrument state
1194 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)