3 Short-Range Comms & IoT Mode
3.3 LoRa (CSS) Demodulation Measurement
Auto Initialize
When the auto initialize function is true for the specified marker, the marker’s trace
is pre-determined automatically by the analyzer whenever the marker is on (Normal,
Delta or Fixed).
When the parameter is switched off for the specified marker, the association
between the marker and its trace remains.
This state of a marker is switched off automatically whenever a trace is manually
designated to the marker.
Remote
Command
:CALCulate:LORA:MARKer[1]|2|...|12:TRACe:AUTO OFF | ON | 0 | 1
:CALCulate:LORA:MARKer[1]|2|...|12:TRACe:AUTO?
Couplings Auto initialize is not impacted by the Auto Couple function
Auto initialize is set to true on a preset or when all markers off is executed
Preset
ON
More Information
When the marker moves between traces the marker’s X position in trace points is
retained as it moves. For moving between active traces this generally means the x-
axis value of the marker will not change. But for moving to or from an inactive trace,
the x-axis value will take on that of the new trace at the bucket the marker was on
the old trace (and is still on, on the new trace, since the bucket does not change).
Note that this is true even if the marker is off screen. Thus, a marker that is at the
center of the screen on the old trace stays at the center of the screen on the new
trace. A marker that is off screen one whole screen to the left on the old trace
remains off screen one whole screen to the left on the new trace – even if this means
it will be at negative time.
Auto Init Rules Flowchart
The following flowchart depicts the Auto Init rules:
396 Short Range Comms & IoT Mode User's &Programmer's Reference