9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Questionable Integrity Event Query
Returns the decimal value of the sum of the bits in the Questionable Integrity Event
register.
NOTE
The register requires that the associated PTRansition or NTRansition filters be
set before a condition register bit can set a bit in the event register. The data in
this register is latched until it is queried. Once queried, the register is cleared.
Remote Command
:STATus:QUEStionable:INTegrity[:EVENt]?
Example
:STAT:QUES:INT?
Preset 0
Status Bits/OPC depend-
encies
Sequential command
Questionable Integrity Negative Transition
Determines which bits in the "Questionable Integrity Condition" on page 2499
register will set the corresponding bit in the "Questionable Integrity Event Query" on
page 2500 register when the condition register bit has a negative transition (1 to 0).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:INTegrity:NTRansition <integer>
:STATus:QUEStionable:INTegrity:NTRansition?
Example
:STAT:QUES:INT:NTR 8
Uncalibrated Summary being cleared will be reported to the Integrity Summary of the Status
Questionable register
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Questionable Integrity Positive Transition
Determines which bits in the "Questionable Integrity Condition" on page 2499
register will set the corresponding bit in the "Questionable Integrity Event Query" on
page 2500 register when the condition register bit has a positive transition (0 to 1).
Short Range Comms & IoT Mode User's &Programmer's Reference 2500