6 Input/Output
6.1 RF Source
Time or Count <time/int> - specifies time duration in seconds or play count of the ARB file associated
with the step
If input parameter number exceeds the step number defined by "Number of Steps" on page 1934,
then an error is generated, and only those parameters whose index number falls within number of
steps will be updated
If current "Step Duration" on page 1944 is “Continuous”, then error -221, "Settings conflict; Cannot
accept time or count input when step duration type is Continuous on step #" is generated
Dependencies The range is 1 to 1000, which is determined by the number of steps you have configured. For details
see "Number of Steps" on page 1934
Step Configuration of Output Trigger parameter list (Remote Command Only)
Configures the Output Trigger parameter array of the whole List Sequencer at one
time. The number of arrays is same as step number defined in "Number of Steps" on
page 1934. As a step is set up, the value entered runs through several levels of
validation.
Remote Command
:SOURce:LIST:SETup:OUTPut:TRIGger <bool>, <bool>, <bool>,
:SOURce:LIST:SETup:OUTPut:TRIGger?
Example
:SOUR:LIST:SET:OUTP:TRIG ON,OFF,ON
:SOUR:LIST:SET:OUTP:TRIG?
Notes Sets up or queries the parameter array of whole list sequence
Output Trigger <Boolean> - specifies the output trigger for the step. Options are:
ON|OFF|1|0
If input parameter number exceeds the step number defined by "Number of Steps" on page 1934,
then error -221,"Settings conflict; The number of input parameters is too large and is truncated to
current list step number" is generated, and only those parameters whose index number falls within
legal step number are updated
Dependencies The range is 1 to 1000, which is determined by the number of steps you have configured. For details
see "Number of Steps" on page 1934
Clear List (Remote Command Only)
The SCPI equivalent of the Clear List UI feature described in "Clear List" on page
1935.
Remote Command
:SOURce:LIST:SETup:CLEar
Example
:SOUR:LIST:SETup:CLE
Dependencies Not available in E7760B
1955 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals