8 Trigger
8.1 Trigger
Notes Not all measurements have all the trigger sources available to them. Check the trigger source
documentation for your specific measurement to see what sources are available
Note that not all trigger sources are available for each input, and that the available parameters are
model number and hardware dependent
For the I/Q Trigger Source, the following trigger sources are available:
IMMediate
free run triggering
EXTernal1
or EXTernal
triggers on an externally connected trigger source on the rear panel
EXTernal2
triggers on an externally connected trigger source on the front panel
IQMag
triggers on the magnitude of the I/Q signal
IDEMod
triggers on the I/Q signal's demodulated I voltage
QDEMod
triggers on the I/Q signal's demodulated Q voltage
IINPut
triggers on the I channel's ADC voltage
QINPut
triggers on the Q channel's ADC voltage
AIQMag
triggers on the magnitude of the auxiliary receiver channel I/Q signal
*OPC should be used after requesting data. This will hold off any subsequent changes to the selected
trigger source, until after the sweep is completed and the data is returned
Available ranges, and from mode-to-mode presets can vary
Status Bits/OPC
dependencies
The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or
Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event
occurs and all the applicable trigger criteria have been met). A corresponding pop-up message
("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec. This message
goes away when a trigger signal appears
8.1.1.1 Free Run
Free Run triggering occurs immediately after the sweep/measurement is initiated.
Example Swept SA measurement:
:TRIG:SOUR IMM
Measurements other than Swept SA:
:TRIG:<meas>:SOUR IMM
Annunciation Free Run (in the Meas Bar)
8.1.1.2 Video/ADC
The Video trigger condition is met when the video signal at the left edge of the
graticule (the filtered and detected version of the input signal, including both RBW
and VBW filtering) crosses the video trigger level with the chosen slope.
Short Range Comms & IoT Mode User's &Programmer's Reference 2328

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals