3 Short-Range Comms & IoT Mode
3.4 HRP UWB Demodulation Measurement
Constraint Length
Sets the convolutional code to use. CL3 is the original code defined in the 2015
standard. CL7 is defined in the 802.15.4z standard.
Remote
Command
[:SENSe]:HUWB:DATA:CONStraint:LENGth CL3 | CL7
[:SENSe]:HUWB:DATA:CONStraint:LENGth?
Example
:HUWB:DATA:CONS:LENG CL3
Couplings Only available when "PHY Mode" on page 501 is ERDEV-HPRF
Preset
CL3
State Saved Saved in instrument state
Range
CL3 | CL7
Data Rate
Sets the ERDEV data rate.
Remote Command
[:SENSe]:HUWB:DATA:RATE LOW | HIGH
[:SENSe]:HUWB:DATA:RATE?
Example
:HUWB:PHR:DATA:RATE LOW
Couplings Not available when "PHY Mode" on page 501 is Non-ERDEV
Preset
LOW
State Saved Saved in instrument state
Range
LOW|HIGH
Frame Length (Octets)
Sets the length of PHY payload field in octets.
Remote
Command
[:SENSe]:HUWB:DATA:PSDU:LENGth <integer>
[:SENSe]:HUWB:DATA:PSDU:LENGth?
Example
:HUWB:DATA:PSDU:LENG 8
:HUWB:DATA:PSDU:LENG?
Preset 33
State Saved Saved in instrument state
Min 0
Max 127
512 Short Range Comms & IoT Mode User's &Programmer's Reference