3 Short-Range Comms & IoT Mode
3.12 Monitor Spectrum Measurement
3.12.8 Meas Setup
This menu panel contains functions for setting up the measurement parameters and
also contains functions for setting up parameters global to all measurements in the
Mode.
3.12.8.1 Settings
Contains frequently used Meas Setup functions to which you will want the fastest
access.
Avg|Hold Num
Specifies the number of measurement averages used when calculating the
measurement result. The average is displayed at the end of each sweep.
After the specified number of average counts, the averaging mode (terminal control)
setting determines the averaging action.
Remote
Command
[:SENSe]:MONitor:AVERage:COUNt <integer>
[:SENSe]:MONitor:AVERage:COUNt?
Example
:MON:AVER:COUN 25
:MON:AVER:COUN?
Preset 10
State Saved Yes
Min/Max 1/1000
Annotation The average count is displayed in the measurement bar on the front panel display. The annotation
appears in the format n/N where n is the current average and N is the average count
Averaging On/Off
Turns averaging on or off.
Remote Command
[:SENSe]:MONitor:AVERage[:STATe] OFF | ON | 0 | 1
[:SENSe]:MONitor:AVERage[:STATe]?
Example
:MON:AVER ON
:MON:AVER?
Preset
OFF
State Saved Yes
Range
OFF|ON
Short Range Comms & IoT Mode User's &Programmer's Reference 1707