6 Input/Output
6.5 Corrections
Examples
The units that may be specified and what appears in the file and on the screen are
shown below:
Transducer
Unit
SCPI Example In the
Correction file
On the screen (also Y Axis Unit
forced to)
dBµV/m :CORR:CSET:ANT
UVM
Antenna
Unit=µV/m
dBµV/m
dBμA/m :CORR:CSET:ANT
UVA
Antenna
Unit=µA/m
dBμA/m
dBμA :CORR:CSET:ANT UA Antenna Unit=µA dBμA
dBpT :CORR:CSET:ANT
PTES
Antenna
Unit=pTesla
dBpT
dBG :CORR:CSET:ANT
GAUS
Antenna
Unit=Gauss
dBG
None :CORR:CSET:ANT
NOC
Antenna Unit=
(or no line at all)
none (not forced)
6.5.6.4 Description
Sets an ASCII description field which will be stored in an exported file. Can be
displayed in the active function area by selecting as the active function, if desired to
appear in a screen capture.
Remote Command
[:SENSe]:CORRection:CSET[1]|2|…|16:DESCription "text"
[:SENSe]:CORRection:CSET[1]|2|…|16:DESCription?
Example
:CORR:CSET1:DESC "11941A Antenna correction"
Notes 45 chars max; may not fit on display if max chars used
Preset Unaffected by a Preset. Set to empty by Restore Input/Output Defaults
State Saved Saved in instrument state
6.5.6.5 Comment
Sets an ASCII comment field which will be stored in an exported file. Can be
displayed in the active function area by selecting as the active function, if desired to
appear in a screen capture.
Remote Command
[:SENSe]:CORRection:CSET[1]|2|…|16:COMMent "text"
[:SENSe]:CORRection:CSET[1]|2|…|16:COMMent?
2125 Short Range Comms & IoT Mode User's &Programmer's Reference