3 Short-Range Comms & IoT Mode
3.8 SEM Measurement
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:FREQuency:STOP?
Example
:SEM:OFFS:INN:LIST:FREQ:STOP 2.715 MHz, 3.515 MHz, 4.00 MHz, 8.00 MHz, 12.50
MHz, 15.0 MHz
:SEM:OFFS:INN:LIST:FREQ:STOP?
Notes Comma-separated list of values
OFFSet 1 is for BTS, 2 for MS. Default is BTS
If the offset is outside the frequency range, the result spectrum will be invalid
Couplings Coupled to "Start Freq" on page 1142. If Stop Freq is lower than Start Freq, Start Freq is automatically
adjusted to (Stop Freq - 100Hz)
Preset
Mode Values
MSR 215 kHz, 1.015 MHz, 1.5 MHz, 10.5 MHz, 50 MHz, 50 MHz, 50 MHz, 50 MHz, 50
MHz, 50 MHz, 50 MHz, 50 MHz
5G NR 5.05 MHz, 10.05 MHz, 40 MHz, 100 MHz, 500 MHz, 500 MHz, 500 MHz, 500
MHz, 500 MHz, 500 MHz, 500 MHz, 500 MHz | 985.0 kHz, 4.50 MHz, 99.500
MHz, 104.5 MHz, 500 MHz, 500 MHz, 500 MHz, 500 MHz, 500 MHz, 500 MHz,
500 MHz, 500 MHz
LTEAFDD,
LTEATDD
5.05 MHz, 10.05 MHz, 15 MHz, 30 MHz, 40 MHz, 50 MHz, 50 MHz, 50 MHz, 50
MHz, 50 MHz, 50 MHz, 50 MHz | 985.0 kHz, 4.50 MHz, 5.5001 MHz, 9.50
MHz,20 MHz, 40 MHz , 40 MHz, 40 MHz, 40 MHz, 40 MHz, 40 MHz, 40 MHz
State Saved Saved in instrument state
Min/Max 100 Hz/Depends on instrument maximum frequency. Same as the Max Span on Swept SA
Measurement
Res BW
Specifies which Resolution BW filter to use when measuring the currently selected
offset.
Offset Res BW Mode allows the instrument to determine the optimum Resolution
BW filter to use when measuring the currently selected offset. using front panel and
all the offsets using SCPI. When changing the Meas BW parameter, if the Res BW
needs to be changed to adhere to the rule:
(N x Res BW) <= (Stop freq of the offset - Start freq of the offset),
where N is the multiplier, this setting will automatically be changed to manual.
When sending the remote command, missing values are not permitted; that is, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values
remain unchanged. The query for this parameter always returns 12 values.
Remote
Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:BANDwidth[:RESolution] <bandwidth>,
1144 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals