3 Short-Range Comms & IoT Mode
3.4 HRP UWB Demodulation Measurement
Abort (Remote Command Only)
Stops the current measurement. Aborts the current measurement as quickly as
possible, resets the sweep and trigger systems, and puts the measurement into an
"idle" state. If the instrument is in the process of aligning when :ABORt is sent, the
alignment finishes before the abort function is performed, so :ABORt does not abort
an alignment.
If the instrument is set for Continuous measurement, it sets up the measurement
and initiates a new data measurement sequence with a new data acquisition
(sweep) taken once the trigger condition is met.
If the instrument is set for Single measurement, it remains in the "idle" state until an
:INIT:IMM command is received.
Remote Command
:ABORt
Example
:ABOR
Notes If :INIT:CONT is ON, then a new continuous measurement will start immediately, with sweep (data
acquisition) occurring once the trigger condition has been met
If :INIT:CONT is OFF, then :INIT:IMM is used to start a single measurement, with sweep (data
acquisition) occurring once the trigger condition has been met
Dependencies For continuous measurement, :ABORt is equivalent to the Restart key
Not all measurements support this command
Status Bits/OPC
dependencies
The STATus:OPERation register bits 0 through 8 are cleared , except bit 6 (Waiting for Periodic
Sync). Initiating a measurement and Waiting for Periodic Sync could be asynchronous
The STATus:QUEStionable register bit 9 (INTegrity sum) is cleared
Since all the bits that feed into OPC are cleared by :ABORt, the Abort command will cause the *OPC
query to return true
X Scale
Accesses controls that let you set the horizontal scale parameters.
Width
Set the width of the X axis which is displayed for the selected trace. The X width can
be set less than the span for frequency-domain traces, enabling you to zoom in on
just a portion of the measured values. Likewise, it can be less than time span
covered by time-domain data. This plus the X Reference Value and X Reference
Position control the range of X values that can be displayed on a trace.
Remote Command
:DISPlay:HUWB:WINDow[1]|2|3|4|8:X[:SCALe]:WIDTh <real>
:DISPlay:HUWB:WINDow[1]|2|3|4|8:X[:SCALe]:WIDTh?
534 Short Range Comms & IoT Mode User's &Programmer's Reference